SIT CENTRE No. 124
Metrological laboratory |
|
 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
RELATIVE HUMIDITY MEASUREMENTS
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
Instruments to be calibrated by Sit Centre no.124:
-
electrical hygrometers and thermohygrometers (*)
-
thermohygrometers transmitters (*)
mechanical hygrographs and thermohygrographs (*)
electric psychrometers (*) (measuring range from 10% RH to 92% RH at air temperature from 0°C to 60°C)
saturated salt solutions (measuring range from 10% RH to 90% RH at the air temperature of 23°C)
Condensing mirror hygrometers (measuring range from -20°C to 60°C)
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
The SIT certificate has to be requested at the time of placing the order, otherwise it's necessary to send the instrument back to the Centre for certification.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
On request, and only if necessary details are available, the instrument setting can be carried out before certification; the relevant charge has to be previously stated.
|
|
|
|
|
|
|
|
|
|
|
|
 |
At the time of placing the order it's necessary to specify the kind of Isotherm requested.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
Minimum points, for the requested Isotherm, are those suggested by the Primary Institute.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
Certification will be according to the available equipment.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
All the instruments of other manufacturers sent for certification have to be provided with their instructions manual.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
UNCERTAINTY WITH WHICH THE SIT CENTRE WAS APPROVED FOR RELATIVE HUMIDITY
|
|
|
|
|
|
|
|
|
|
|
MAGNITUDE
|
INSTRUMENTS TO BE CALIBRATED
|
MEASURING RANGE
|
UNCERTAINTY(*)
|
| relative humidity |
Electric and mechanical hygrometers and thermohygrometers
Electric psychrometers
|
from 10%RH to 92%RH (at air temperature from 0° to 60°C)
from 10%RH to 92%RH (at air temperature from 0° to 60°C)
|
from 0.5%RH. a 1.8%RH.
from 0.5%RH. a 1.8%RH.
|
| relative humidity |
Saturated salt solutions
|
from 10%RH to 90%RH (at air temperature from 20° to 25°C)
|
1.4%RH
|
| dew point |
Condensing mirror hygrometers
|
from -20°C to 60°C
|
0.16°C
|
|
|
|
|
|
|
|
|
|
(*) The measuring uncertainties are express as twice the standard deviation, corresponding in the case of normal distribution to a confidence level of about 95%.
|
|
|
|
|
|
 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Delta OHM - Via G.Marconi,5 35030-Caselle di Selvazzano (PD)
** * -- * --
|
|
|
|
|